Model Number 71355094
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/26/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that before surgery, while setting up for the case, it was difficult to connect the reamer quick connect handle to the distal reamer, a second handle (back up) was used but the same issue was found with this instrument.After a gentle tap of the handles on the edge of the tray to help loosen the shaft, large shards of dried blood fell from the device into the tray of reamers (tray was deemed unsterile).A competitor¿s device was used to complete the surgery, no patient injury was reported.
 
Additional Manufacturer Narrative
The devices, intended for use in treatment, were not returned for evaluation.A visual inspection of the provided pictures confirmed the stated failure mode.The devices had visible biological contamination on the connection end, after sliding back the locking sleeve.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device passed clean ability testing in 2011 and should therefore be able to be cleaned per the validated cleaning process outlined in instructions for care, maintenance, cleaning and sterilization of smith and nephew orthopaedics devices.These devices are considered devices with challenging design features in our cleaning instructions as they may have retractable and moving features.The potential probable cause of this event is likely inadequate cleaning.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the devices or additional information be received, the complaints will be reopened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER QUICK CONNECT
Common Device Name
PRSTH,HP,SM-CNSTR,MTL/CRM/PLY,CMN R N-PRS,UNCMN
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9580283
Report Number1020279-2020-00226
Device Sequence Number11574306
Product Code LZO
UDI-Device Identifier00885556223093
UDI-Public00885556223093
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113789
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71355094
Device Catalogue Number71355094
Device Lot Number17GSC0001
Initial Date Received by Manufacturer 12/26/2019
Supplement Date Received by Manufacturer04/02/2020
Initial Report FDA Received Date01/13/2020
Supplement Report FDA Received Date04/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-