BOSTON SCIENTIFIC CORPORATION GUIDEZILLA II; CATHETER, PERCUTANEOUS Back to Search Results
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2019
Event Type  malfunction  
Manufacturer Narrative
Initial reporter city: (b)(6).
 
Event Description
It was reported that the balloon shaft was separated during removal.A 10mmx3.50mm wolverine coronary cutting balloon and guidezilla ii guide extension catheter were selected for use.During withdrawal, it was noted that the balloon catheter shaft became separated as it was not compatible with the guidezilla ii.The device was completely pulled out from the patient's body.The device was still intact and there were no fragments left inside the patient.The procedure was completed with another device.No patient complications were reported and the patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA II
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9580634
MDR Text Key177408938
Report Number2134265-2020-00072
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/23/2019
Initial Date FDA Received01/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-