DORNOCH DUO FLUID CART; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number 00514010100
Device Problem Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Date 12/18/2019
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded under zimmer biomet complaint number (b)(4).The device has been evaluated by an external contractor and the investigation is in process.Once the investigation is completed, a supplemental report will be filed accordingly.
 
Event Description
It has been reported that the cart had a burning smell.The event timing was during cleaning.There was no patient involvement.No adverse events have been reported as a result of this malfunction.
 
Event Description
No additional event information.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Udi : (b)(4).On 19 december 2019, a zimmer biomet certified service repair technician was contacted about the cart and dispatched to be at the site.The technician arrived at the site but could not find any issue with the cart.The technician verified that the unit was functioning as intended then returned the unit to service without further incident.The reported event could not be confirmed during inspection of the device; therefore, the root cause for the unit having a burning smell could not be determined.The device was noted to be functioning as intended and no problem was found.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Review of the information provided during the investigation determined there is no further actions needed at this time.This complaint will be tracked and trended per complaint trending procedure for any adverse trends that may warrant further action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
MDR Report Key9581207
MDR Text Key194601401
Report Number0001954182-2020-00002
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00514010100
Device Lot Number0024489
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-