| Model Number |
63501 |
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Explant analysis showed silicone debris in the posterior valve channel, inhibiting seal of the valve and causing a leak.
|
| |
|
Event or Problem Description
|
|
Deflation resulting in explantation.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant analysis showed silicone debris in the posterior valve channel, inhibiting seal of the valve and causing a leak.
|
| |
|
Event or Problem Description
|
|
Deflation resulting in explantation.
|
| |
|