Model Number 1012449-08
Medical Device Problem Codes Inflation Problem (1310); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was performed to treat a mildly calcified, mildly tortuous de novo left anterior descending artery.The 3.0x8mm nc trek balloon dilatation catheter was advanced to post-dilate a 3.0x23mm xience xpedition stent however, the nc trek balloon failed to inflate and a leak was noted from the shaft.Therefore, the device was removed and replaced with a new same size nc trek to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual, functional and sem inspections/analysis were performed on the returned device.The reported tear and inflation issue were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation determined the noted longitudinal tear in the inner member, leak and inflation issue appear to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9582231
Report Number2024168-2020-00611
Device Sequence Number16277342
Product Code LOX
UDI-Device Identifier08717648151859
UDI-Public08717648151859
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model Number1012449-08
Device Catalogue Number1012449-08
Device Lot Number90710G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2019
Initial Date Received by Manufacturer 12/23/2019
Supplement Date Received by Manufacturer03/04/2020
Initial Report FDA Received Date01/13/2020
Supplement Report FDA Received Date03/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.0X23MM XIENCE XPEDITION
-
-