| Model Number |
0600310 |
| Medical Device Problem Code |
Material Perforation (2205)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample was returned for evaluation.The company is still investigating the issue at this time.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model 0600310.Chronic catheter allegedly experienced material perforation.This report was received from a single source.This event did involve patient with no reported patient injury.The patients age, weight and gender were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot number for the malfunction is unknown, therefore the manufacturing review could not be conducted.The device has been returned for evaluation; the evaluation identified circumferential rupture and break along with leak from the split.Based upon the available information, a definitive root cause is unknown.The device is labeled for single use.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model 0600310 chronic catheter allegedly experienced material perforation.This report was received from a single source.This event did involve patient with no reported patient injury.The patients age, weight and gender were not provided.
|
| |
|
Search Alerts/Recalls
|