Model Number FV314T
Device Problem Infusion or Flow Problem (2964)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 12/12/2019
Event Type  Injury  
Manufacturer Narrative
Height: 94cm.When additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported that valve is under-draining.The reporter indicated that a 2 year 11 month post operative valve under-drains.Additional information has not been provided.
 
Manufacturer Narrative
Investigation: visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: this is a fixed pressure valve.An adjustment test is not applicable.Braking force and brake function test: this is a fixed pressure valve.A braking force and brake function test is not applicable.Computer controlled test: to investigate the claim of underdrainage, the opening pressure is measured using a miethke computer controlled testing apparatus, which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve operates within the accepted tolerance in both positions.Results: first, we performed a visual inspection of the gav.No significant deformations or damage of the valve were detected during the visual inspection.Next, we tested the permeability of the valve.The valve was shown to be permeable.Then we carried out a computer controlled simulated flow test.The measured opening pressure was within the accepted tolerance in both positions.Finally, we have dismantled the valve.Inside the valve we have found a build-up of substances (likely protein).Based on our investigation, we are unable to substantiate the claim of underdrainage.At the time of our investigation, the valve was operating within the specified tolerances.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAV VALVE 10/40
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO. KG
2 ulanenweg
potsdam d, 14469
GM  14469
MDR Report Key9583728
MDR Text Key188558764
Report Number3004721439-2020-00006
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
K031303
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Model NumberFV314T
Device Catalogue NumberFV314T
Device Lot Number20031645
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received01/13/2020
Supplement Dates Manufacturer Received01/10/2020
Supplement Dates FDA Received01/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age4 YR
Patient Weight15
-
-