Investigation: visual inspection: no significant deformations or damage of the valves were detected during the visual inspection.Permeability test: a permeability test has shown that the valves are permeable adjustment test: the adjustment test has shown that the progav 2.0 is adjustable to all settings.For the shunt assistant 2.0 the adjustment test is not applicable, because it is a valve with a fixed pressure.Braking force and brake function test: the investigation of the braking force of the progav 2.0 valve showed that the brake function is fully operational and the braking force is within the given tolerances.For the shunt assistant 2.0 the test is not applicable, because it is a valve with a fixed pressure.Results: first, we performed a visual inspection of the valves.No significant deformations or damages of the valves were detected during the visual inspection.Further, we tested the permeability of the valves.The test has shown that the valves were permeable.The testing of the adjustability, the brake functionality and as well the brake force of the progav 2.0 valve was possible and within the accepted tolerances.For the shunt assistant 2.0 the test is not applicable because it is a valve with a fixed pressure.Finally, we have dismantled the valves.There were no visible deposits observed inside the valves.We cannot confirm the suspected blockage.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.Further actions: no further actions are required in our point of view.
|