ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRM-600
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported right side rupture.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: opening on radius, black particles and underweight.A visual and micro analysis was performed which identified: sharp opening and crease fold.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on radius due to an unidentified (tear) opening.
 
Event or Problem Description
Healthcare professional reported right side rupture.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9583748
Report Number9617229-2019-23320
Device Sequence Number534981
Product Code FTR
UDI-Device Identifier10888628008182
UDI-Public10888628008182
Combination Product (Y/N)N
Initial Reporter StateTX
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/21/2020
Device Model NumberTRM-600
Device Catalogue NumberTRM-600
Device Lot Number2857591
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Initial Date Received by Manufacturer 12/20/2019
Supplement Date Received by Manufacturer01/29/2020
Initial Report FDA Received Date01/13/2020
Supplement Report FDA Received Date02/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-