| Model Number |
ZXT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but best estimate date is between 7/15/2019 and 11/20/2019.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt150 intraocular lens (iol) was implanted in the patient's right (od) eye on (b)(6) 2019.It was explanted on (b)(6) 2019 because of residual myopia.There was no patient injury.No surgical interventions such as: vitrectomy; incision enlargement; or sutures were required.The replacement lens is a zxt150 19.0 diopter lens.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 1/23/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site in a zip lock bag.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint issue reported was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no similar complaints were received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.H3 other text : placeholder.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|