• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number MC2-4590S
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.No root cause can be confirmed at this time.
 
Event or Problem Description
Information was received that x-rays revealed that the end cap had disengaged.At this time it is unknown if the patient will undergo a revision procedure.
 
Additional Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Additional Manufacturer Narrative
Section e1 was updated to reflect the correct facility name and address.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
MDR Report Key9583884
Report Number3006179046-2020-00101
Device Sequence Number3012810
Product Code PGN
UDI-Device Identifier00812258022624
UDI-Public812258022624
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Recall,Notification
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-4590S
Device Catalogue NumberPA0684-001
Device Lot Number8070223
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/17/2020
Supplement Date Received by Manufacturer04/01/2020
09/18/2020
Initial Report FDA Received Date01/13/2020
Supplement Report FDA Received Date04/01/2020
09/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1898-2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-