STRYKER ORTHOPAEDICS-MAHWAH BOX CHISEL; CHISEL, SURGICAL, MANUAL Back to Search Results
Catalog Number 6541-4-709
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 12/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding an intraoperative fracture involving a triathlon chisel was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: could not be performed as lot code information was not provided.Complaint history review: could not be performed as lot code information was not provided.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Not returned to the manufacturer.
 
Event or Problem Description
Primary procedure, left knee.It was reported that after using the ps box guide and chisel, a crack in the patient's medial condyle was noted.The fracture was addressed using a k-wire, and the surgery was completed with no delay.The rep reported that the devices will be kept in use by the hospital, and that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOX CHISEL
Common Device Name
CHISEL, SURGICAL, MANUAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9584506
Report Number0002249697-2020-00075
Device Sequence Number14206480
Product Code FZO
UDI-Device Identifier07613327179347
UDI-Public07613327179347
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K123486
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number6541-4-709
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/16/2019
Initial Report FDA Received Date01/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient Weight95
-
-