| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had blank display.The customer¿s blood glucose level was 138 mg/dl at the time of the incident.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer stated the display was not blank less than 30 seconds and did not return.Advised customer to remove the battery and customer stated the insert battery alarm did not display on the insulin pump screen.Customer stated the contact on the battery cap was not missing or damage or corroded.Customer stated battery compartment was neither damaged nor corroded.Customer stated the spring was neither damaged nor corroded.Customer stated the display did not return after insulin pump restart.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The customer was advised the insulin pump will be replaced as per troubleshoot.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device monitored for several days and no blank display noted.Device received with normal operating current.Device passed self test.Device passed off no power test.
|
| |
|
Search Alerts/Recalls
|