MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SRLP-220
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 01/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade unknown.Device remains implanted.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b.5, d.7, d.10, g.1, g.3, h.3, h.6.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified crease flat, deformation, and cloudiness in the gel observed after autoclave disinfection process.Weight measurement of the device identified device weight was within specification.Based on the device analysis the final assessment was no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade unknown.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9585802
Report Number9617229-2019-23147
Device Sequence Number14300085
Product Code FTR
UDI-Device Identifier10888628006850
UDI-Public10888628006850
Combination Product (Y/N)N
Initial Reporter StateCA
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/16/2022
Device Model NumberSRLP-220
Device Catalogue NumberSRLP-220
Device Lot Number3033307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2020
Initial Date Received by Manufacturer 12/19/2019
Supplement Date Received by Manufacturer02/24/2020
04/02/2020
Initial Report FDA Received Date01/14/2020
Supplement Report FDA Received Date03/24/2020
04/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-