Model Number 030450
Medical Device Problem Codes Component or Accessory Incompatibility (2897); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, pre-operatively, when the reload was installed, the jaw could not close.There was no patient involvement.
 
Additional Manufacturer Narrative
Additional information: b5, d10, g4, h3, h6 h3.Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual examination noted that the loading unit had a full complement of staples and the jaws were open.The loading unit was loaded into a pmv representative instrument for functional testing.The loading unit was applied to test media.All staples were placed and the test media was cleanly transected.Functional testing confirmed the safety interlock feature successfully prevented the loading unit from cycling again.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use, on a nephrectomy, when the reload was installed, the jaw could not close and the device could not be loaded.They changed the product to a different manufacturer to complete the case.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9586102
Report Number1219930-2020-00292
Device Sequence Number19680339
Product Code GDW
UDI-Device Identifier10884523002812
UDI-Public10884523002812
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number030450
Device Catalogue Number030450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2020
Initial Date Received by Manufacturer 12/31/2019
Supplement Date Received by Manufacturer05/08/2020
Initial Report FDA Received Date01/14/2020
Supplement Report FDA Received Date05/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-