| Model Number |
FEM14080 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Entrapment of Device (1212); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device for this malfunction has been returned to the manufacturer for evaluation.The investigation of the reported malfunction is currently underway.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model fem14080 endovascular stent graft allegedly experienced a difficulty to deploy and malposition of device.This information was received from a single source.The malfunction involved a patient with no reported consequence.The patient was reported as a (b)(6)-year-old female whose weight was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the lot number for the malfunction was provided and a lot history review was performed.The device has been returned for evaluation; the evaluation confirmed difficult to deploy, however, unconfirmed malposition and entrapment of device.Further evaluation of the device, added a entrapment device code (1212).Based upon the available information, a definitive root cause is unknown.The device is labeled for single use.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model fem14080 endovascular stent graft allegedly experienced a difficulty to deploy and malposition of device.This information was received from a single source.The malfunction involved a patient with no reported consequence.The patient was reported as a 73-year-old female whose weight was not provided.
|
| |
|
Search Alerts/Recalls
|