Model Number DM3500
Device Problem Signal Artifact/Noise (1036)
Patient Problem Syncope (1610)
Event Date 01/10/2020
Event Type  malfunction  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient experienced an episode of syncope.Upon interrogation, the device displayed no episodes of syncope due to suspected noise.A device upgrade procedure was performed which was unrelated to the event and the patient was in stable condition.
 
Manufacturer Narrative
Analysis performed indicated the device exhibited normal characteristics.A sensitivity test was performed on the device with various settings and it was able to sense within the product specification.Visual inspection did not find any foreign material on the header feed through pin that could contribute to the noise complaint.A longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.The reported event of noise was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9586385
MDR Text Key175706308
Report Number2017865-2020-00397
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/18/2020
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000069523
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-