• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TRAUMAONE SYSTEM DRILL, 1.6X50MM 22MM STOP; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TRAUMAONE SYSTEM DRILL, 1.6X50MM 22MM STOP; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore a facility medwatch report will not be available.Report source ¿ (b)(6).
 
Event or Problem Description
It was reported a drill bit fractured while drilling a pilot hole during surgery for a mandibular fracture.The fractured piece of the instrument was removed from the patient.No additional patient consequences were reported.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The drill was returned for evaluation.On visual inspection, the drill was noted to have a transverse fracture through the flutes near the neck.The cert was reviewed and there were no non-conformances were found.There are no indications of manufacturing defects.For this part (46-1006) and the previous one year (from the notification date) regarding the drill fracturing, there is a complaint rate of 0.19%, which is no greater than the occurrence listed in the application fmea.This is the only complaint in regards to the drill fracturing for 46-1006 serial number (b)(6).The most likely underlying cause of the complaint is the drill experienced force in excess of what it was designed to encounter; fractures at this point of the drill are typical of excessive force fractures.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAUMAONE SYSTEM DRILL, 1.6X50MM 22MM STOP
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9586527
Report Number0001032347-2020-00059
Device Sequence Number11628372
Product Code HBE
UDI-Device Identifier00841036137410
UDI-Public00841036137410
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number46-1006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/17/2019
Supplement Date Received by Manufacturer03/20/2020
Initial Report FDA Received Date01/14/2020
Supplement Report FDA Received Date04/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-