U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CYSTOSCOPY IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number 2C4040
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 12/16/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported nineteen (19) cysto/bladder irrigation sets were contaminated.It was further reported the tubing appeared to have "diffuse/random greyish-black discoloration that transverses the tubing laterally (when cut open for inspection)." there was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added.Five (5) devices were received for evaluation.One (1) of the five samples was received having a cut on the tubing.A visual inspection was performed using the naked eye which observed a brown/yellow discoloration between the walls of the tubing, which caused a visual effect of contamination in the tubing.The discoloration was cosmetic in nature and no particulate matter (pm) was observed in the fluid path of the tubing.Therefore, the reported condition of pm in fluid path was not verified.Functional testing was not performed for this complaint.The other samples were not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTOSCOPY IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9587715
MDR Text Key175027435
Report Number1416980-2019-07557
Device Sequence Number1
Product Code LJH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960787
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C4040
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2019
Initial Date FDA Received01/14/2020
Supplement Dates Manufacturer Received01/27/2020
Supplement Dates FDA Received02/17/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-