Model Number 283489
Medical Device Problem Codes Mechanical Jam (2983); Suction Failure (4039)
Health Effect - Clinical Codes Edema (1820); Not Applicable (3189)
Date of Event 01/08/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event or Problem Description
It was reported by affiliate via email, that during an arthroscopic shoulder surgery (subacromial decompression, clavicle resection) the shaver blad started to sound strange, had bad suction and stopped working.When the blade was taken out of the handpiece, the inner and outer parts were stuck together and could not be separated.The blade was replaced with a similar one, but the procedure was prolonged 15 min.The surgeon experienced a bit more swelling than normal.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that there was no problems with the fms vue.Only the shaver blade had the issue.
 
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: investigation summary
=
> according to the information provided, it was reported that the barrel tornado, started to sound strange, had bad suction and stopped working.The complaint device was received and evaluated.Visual observations confirm that burr inner blade was jammed inside the outer shaft and was corroded, functional testing,was performed, once connected at pump, was observed that the device was jammed and made strange sounds, the complaint can be confirmed.The possible root cause for the reported failure can be occur if the surgeon does not utilize proper irrigation during use of the device.When the proper irrigation is not utilized, the blade surface can heat up from the friction between the inner blade and outer sheath therefore causing a cold-type weld to occur.However, this cannot be conclusive determined.A manufacturing record evaluation was performed for the finished device [m1907056] number, and no non-conformances were identified.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
> a manufacturing record evaluation was performed for the finished device [m1907056] number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARREL TORNADO BURR 4.0MM 5PK
Common Device Name
 ARTHROSCOPIC SHAVER SYSTEM BLADE, SINGLE-USE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9588543
Report Number1221934-2020-00196
Device Sequence Number19680327
Product Code HRX
UDI-Device Identifier10886705022052
UDI-Public10886705022052
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K041824
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model Number283489
Device Catalogue Number283489
Device Lot NumberM1907056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Initial Date Received by Manufacturer 01/08/2020
Supplement Date Received by Manufacturer01/28/2020
02/14/2020
03/15/2020
Initial Report FDA Received Date01/14/2020
Supplement Report FDA Received Date01/29/2020
02/18/2020
03/16/2020
Patient Sequence Number1
-
-