Additional Manufacturer Narrative
|
|
A medtronic representative went to the site to test the equipment.It was reported that the issue was confirmed.The issue was isolated to the malleable suction and using another suction restored functionality.The system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used for a tumor resection ( transsphenoidal) procedure.The site was unable to track the malleable suction during the procedure.The procedure was completed with the use of a second suction.The replacement suction tracked without issue.The issue resulted in less than one hour procedure delay.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: hardware analysis of the returned malleable suction confirmed the allegation.The instrument would recognize when connected to a known good system, but would not track (only show red status).It was determined that coil#1 and coil#3 were dead.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|