• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS PRESTIGE CERVICAL DISC SYSTEM; PROSTHESIS, INTERVERTEBRAL DISC Back to Search Results
Catalog Number 6961460
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Outcomes to adverse event - other - hoarseness; problems swallowing; lack of muscle strength in throat; neck spasms; pain in shoulders and upper back; painful breathing; digestive problems; scoliosis; kyphosis; and dysphagia.Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent cervical surgery to repair disc bulge in her neck.Post-op, the patient claimed that about 4 years ago she had a fall with a concussion- she had also been involved in two previous auto accidents (rear-ended).Patient is experiencing following problems for about 3 years: hoarseness; problems swallowing; lack of muscle strength in throat; neck spasms; pain in shoulders and upper back; painful breathing; digestive problems.Ct scan finally showed the implanted hardware (disc) in her neck had shifted.She has over time developed scoliosis, kyphosis, and dysphagia.One doctor told her it was inoperable and couldn't be repaired.She then got a second opinion and was told it was operable and her vertebrae needed to be fused at 2 levels, but she is having trouble getting someone to do the surgery.She also claims that doctor initially suggested removal, but now his mind would changed.She was searching for another neurosurgeon to give her an opinion.
 
Additional Manufacturer Narrative
X-ray review results: post-op, x-rays for c5-6 prestige artificial disc replacement.There is a mild anterior back out of the inferior screw.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE CERVICAL DISC SYSTEM
Common Device Name
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key9590162
Report Number1030489-2020-00065
Device Sequence Number667322
Product Code MJO
UDI-Device Identifier00613994129215
UDI-Public00613994129215
Combination Product (Y/N)N
PMA/510(K) Number
P060018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6961460
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/17/2019
Supplement Date Received by Manufacturer01/21/2020
Initial Report FDA Received Date01/15/2020
Supplement Report FDA Received Date02/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-