| Model Number |
C12787 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
While performing a ffr case, a burning smell was noted and the power adapter connected to the quantien became burnt.Smoke was noted to be coming from the power adapter and the system shut down.The procedure was completed without ffr with no adverse consequences to the patient or user.The quantien was replaced to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received and electronic data, the reported incident is verified to have occurred, but the cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
One quantien was received into the lab for analysis with no accessories or original packaging available for inspection.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott and the investigation performed, root cause of the field reported event citing a thermal event was successfully isolated to abnormal functionality of the attached power supply.
|
| |
|