• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U354054
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the malfunction was provided and a lot history review will be performed.The device has been returned for evaluation; the evaluation identified material split, cut, or torn.Based upon the available information, a definitive root cause is unknown.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model u354054 pta balloon dilatation catheter allegedly experienced an inflation issue and material rupture.This information was received from one source.The malfunction involved one patient with no patient consequences.The age, weight, and gender of the patient were not provided.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model u354054 pta balloon dilatation catheter allegedly experienced an inflation issue and material split, cut, or torn.This information was received from one source.The malfunction involved one patient with no patient consequences.The age, weight, and gender of the patient were not provided.
 
Additional Manufacturer Narrative
H10: the malfunction was reassessed for reportability and determined to be no longer reportable.The lot number for the malfunction was provided and a lot history review will be performed.The device has been returned for evaluation; the evaluation confirmed inflation issue and material split, cut, or torn but is unconfirmed for material rupture.Based upon the available information, a definitive root cause is unknown.The device is labeled for single use.H10: g4 h11: b5; h6 (device) h11 :section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9591280
Report Number2020394-2020-00277
Device Sequence Number491677
Product Code LIT
UDI-Device Identifier00801741092398
UDI-Public(01)00801741092398
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberU354054
Device Catalogue NumberU354054
Device Lot NumberCMDQ0516
Initial Date Received by Manufacturer 12/31/2019
Supplement Date Received by Manufacturer03/31/2020
Initial Report FDA Received Date01/15/2020
Supplement Report FDA Received Date04/03/2020
Patient Sequence Number1
-
-