• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 15 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 15-213
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 08/13/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Event or Problem Description
Healthcare professional reported bilateral ¿capsular contracture, grade iii¿.The device has been explanted.This record is for the right side.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: weight to the spec, crease fold, deformation, wear abrasion.Cloudy and bubbles in the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported bilateral ¿capsular contracture, grade iii¿.The device has been explanted.This record is for the right side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9591614
Report Number9617229-2020-00671
Device Sequence Number11437234
Product Code FTR
UDI-Device Identifier10888628001411
UDI-Public10888628001411
Combination Product (Y/N)N
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/02/2022
Device Model Number15-213
Device Catalogue Number15-213
Device Lot Number3016536
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Initial Date Received by Manufacturer 12/18/2019
Supplement Date Received by Manufacturer03/09/2020
Initial Report FDA Received Date01/15/2020
Supplement Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
-
-