BIOMET MICROFIXATION TRAUMAONE SYSTEM 1.5/0.6MM, L PLATE, 100 DEGREE, 10MM, LEFT; PLATE, BONE
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
10/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.The device will not be returned for analysis as it was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00066.Medical products: traumaone system 1.5/0.6mm, l plate, 100 degree, 10mm, right, part# sp-3210-r, lot# unk, traumaone system 1.5/0.6mm, l plate, 100 degree, 10mm, left, part# sp-3210-l, lot# unk, unknown screws, part# unk, lot# unk.
|
| |
|
Event or Problem Description
|
|
It was reported that plates fractured post-operatively and were removed in a revision.The patient had a bruxing habit that may have contributed to the fracture.The hardware and fractured plates were removed in a revision.No additional patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.As there was a revision surgery, the complaint is confirmed.The two (2) 1.5/.6mm l plt 100d 10mm left (part# sp-3210-l, lot# unk) were not returned for investigation; therefore, no visual inspections or functional testing could be conducted.No scans, x-rays, pictures, of physician's reports were provided.The dhrs and the non-conformance database could not be reviewed for these parts as the lot numbers remain unknown.There are no indications of manufacturing defects.For this part (sp-3210-l) and the previous five years (from the notification date) regarding the fracturing of this plate, there is a complaint rate of 0.92% which is greater than the occurrence listed in the application fmea.Because this is an isolated incident with a product with low sales, the exceeded occurrence rate does not require an update to the risk assessment document.The most likely underlying cause of the complaint is patient activity as it was reported the patient had a bruxing habit.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|