ARTHROCARE CORP. QFIX 1.8 MINI SUTURE ANCHOR DISP KIT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72290125
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/20/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported about a qfix 1.8 mini suture anchor that during a shoulder procedure the drill bit exhibited to be blunt and difficult to drill.The procedure was successfully completed without significant delay using a back-up device in an additional bone hole.There were no patient injuries and no other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10: h6: the device reported, used in treatment, was not returned for evaluation.The relationship between the product and reported incident cannot be established.A review of manufacturing records for the reported lot number 2029113 found no non-conformances or anomalies during manufacturing process related to the reported event.The review of the complaint history for the associated complaint product found no similar events in the last three years for the involved lot.Without the reported product a fully visual and functional evaluation cannot be performed and customer¿s complaint cannot be confirmed.An exact root cause cannot be determined without evaluation of the device; however, factors unrelated to the manufacture or design of the device that could have contributed to the reported event include: improper alignment of the handle.No containment or corrective actions are recommended at this time.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QFIX 1.8 MINI SUTURE ANCHOR DISP KIT
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key9592606
Report Number3006524618-2020-00023
Device Sequence Number18925948
Product Code NBH
UDI-Device Identifier00885556662632
UDI-Public00885556662632
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2022
Device Model Number72290125
Device Catalogue Number72290125
Device Lot Number2029113
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/20/2019
Supplement Date Received by Manufacturer03/09/2020
Initial Report FDA Received Date01/15/2020
Supplement Report FDA Received Date03/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-