| Model Number |
362706 |
| Medical Device Problem Code |
High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|
Event or Problem Description
|
|
It was reported the patient had a very high and fluctuating ventricular threshold upon new lead insertion.The lead was repositioned with no success and finally it was explanted three days after the implantation.A new lead with active fixation was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, the lead under complaint was subjected to an extensive analysis.The analysis revealed a cut into the insulation (approx.26 cm distal to the is-1 connector pin), which most likely resulted from the explantation procedure.A thorough analysis of the lead did not show any deviations which might have led to the reported observation.In particular the values measured during the electrical analysis were within the specifications.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing process.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|
| |
|