• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTIUMUSA IMPLANTIUM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTIUMUSA IMPLANTIUM; DENTAL IMPLANT Back to Search Results
Model Number FX4308MLC
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 05/17/2019
Type of Reportable Event Serious Injury
Event or Problem Description
The dental implant fx4308mlc was removed on (b)(6) 2019 due to failure to osseointegrate 3 months and 5 days after implantation.The patient has no significant medical history.The patient required bone augmentation for the operation.The doctor noted bone resorption during explantation.The patient required another surgical intervention to recover.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTIUM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DENTIUMUSA
6761 katella ave
cypress CA 90630
MDR Report Key9593672
Report Number3005503242-2020-00002
Device Sequence Number18925953
Product Code DZE
UDI-Device IdentifierD755FX4308MLC0
UDI-PublicD755FX4308MLC0
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K041368
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/19/2023
Device Model NumberFX4308MLC
Device Catalogue NumberFX4308MLC
Device Lot NumberI16SA962S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2020
Initial Report FDA Received Date01/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
-
-