It was reported that the arctic sun device displayed 0 flow.There were only three pads in use because the fourth pad would not stick.The nurse disconnected and reconnected the pads using the proper technique and flow remained at 0lpm.The nurse took off the fluid delivery line and ran diagnostics.The flow rate was 2.2lpm, and inlet pressure was -9psi range.The circulation pump was at 70%.The nurse reconnected the pads one by one and the flow dropped to 0.0 again.Ms&s advised the nurse to change pads.Per follow up call, the patient was able to complete therapy once the pads were changed.
|