| Model Number |
CQF7564 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Of the two malfunctions, two lot numbers were provided, and lot history reviews were performed.A device was returned for evaluation for each malfunction; the evaluation did not identify a deflation issue.A root cause has not been determined.The devices were labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes two malfunctions.A review of the reported information indicated that model cqf7564 pta balloon dilatation catheter allegedly experienced deflation issue.This information was received from various sources.Each malfunctions involved a patient with no known impact to the patient.One patient was a (b)(6) year-old female weighing (b)(6) lb and the other patient was a (b)(6) year-old female weighing (b)(6) lb.
|
| |
|