Catalog Number 283512
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter phone number (b)(6).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the unit micro tornado not was working.Per operational and diagnostic, the reported failure was confirmed.  during evaluation the motor was found corroded and did no turn anymore, therefore the motor was replaced.In addition the insulation resistance was out of range, the motor cable replaced.The handcontrol was not responding properly and the preventive replacement of o-rings was performed.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The possible root cause for the reported failure was thus identified as faulty parts.  as a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.  at this time, no corrective action is required, and no further action is warranted, as the device was repaired and is fully functional.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.Udi:(b)(4).
 
Event Description
It was reported by an affiliate that the micro tornado hp w handcontrol no longer functions.Additional information pertaining to the procedure, patient, or user could not be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key9598121
MDR Text Key207589400
Report Number1221934-2020-00232
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2019
Date Manufacturer Received01/15/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-