MAUDE Adverse Event Report: COLOPLAST A/S TITAN SCROTAL ZERO ANG 18CM; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN SCROTAL ZERO ANG 18CM; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number ES89182400
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 12/18/2019
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information, the device was undersized, resulting in the patient having a very floppy glans.The removed device was measured by the scrub tech as roughly 18 cm.A new 24 cm prosthesis was placed with a 1.5cm rear tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN SCROTAL ZERO ANG 18CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
Manufacturer (Section G)
COLOPLAST MANUFACTURING US, LLC
1601 west river road north
minneapolis MN 55411
Manufacturer Contact
stephanie perryman
1601 west river road north
minneapolis, MN 55411
6124345685
MDR Report Key9599404
MDR Text Key175413163
Report Number2125050-2020-00046
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932487344
UDI-Public05708932487344
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberES89182400
Device Catalogue NumberES8918
Device Lot Number6565101
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2019
Initial Date FDA Received01/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
-
-