• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC SCREWPLANT IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC SCREWPLANT IMPLANT; DENTAL IMPLANT Back to Search Results
Catalog Number 103711P
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 08/23/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Per complaint (b)(4), during laboratory procedure, patient experienced loss or failure of implant to integrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREWPLANT IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
MDR Report Key9600243
Report Number3001617766-2020-00394
Device Sequence Number585522
Product Code DZE
UDI-Device Identifier10841307105077
UDI-Public10841307105077
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K061319
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2019
Device Catalogue Number103711P
Device Lot Number57362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/23/2019
Initial Report FDA Received Date01/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age45 YR
-
-