• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA CUP: MPACT ACETABULAR SHELL Ø50 TWO-HOLES LOT. 161850

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA CUP: MPACT ACETABULAR SHELL Ø50 TWO-HOLES LOT. 161850 Back to Search Results
Catalog Number 01.32.150DH
Health Effect - Clinical Code Joint Disorder (2373)
Date of Event 12/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 20 december 2019.Lot 161850: (b)(4) items manufactured and released on 13-sep-2016.Expiration date: 2021-08-30.No anomalies found related to the problem.To date, all items of the same lot have been already sold without any other similar reported event.Preliminary investigation performed by r&d project manager from preliminary visual investigation it is not possible to determine an implant's related failure root cause.Clinical evaluation: revision surgery performed 3 years after cementless total hip arthroplasty in a (b)(6) man.Radiographic image provided shows a mobilized cup.Initial shell position cannot be determined as no immediate postoperative image is available.Aseptic loosening is a possible literature described adverse event after cementless total hip arthroplasty and causes are often unknown.The reason of this event cannot be determined.
 
Event or Problem Description
Acetabular cup loosening unknown cause, due to mobilization there was an impingement cup- stem.Cup, head and liner were revised almost 3 years and 3 weeks after primary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP: MPACT ACETABULAR SHELL Ø50 TWO-HOLES LOT. 161850
Common Device Name
ACETABULAR SHELL
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
MDR Report Key9601763
Report Number3005180920-2019-01148
Device Sequence Number2040957
Product Code LPH
UDI-Device Identifier07630030810763
UDI-Public07630030810763
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K132879
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/30/2021
Device Catalogue Number01.32.150DH
Device Lot Number161850
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2019
Initial Report FDA Received Date01/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient Weight64
-
-