| Catalog Number |
01.32.150DH |
| Health Effect - Clinical Code |
Joint Disorder (2373)
|
| Date of Event |
12/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 20 december 2019.Lot 161850: (b)(4) items manufactured and released on 13-sep-2016.Expiration date: 2021-08-30.No anomalies found related to the problem.To date, all items of the same lot have been already sold without any other similar reported event.Preliminary investigation performed by r&d project manager from preliminary visual investigation it is not possible to determine an implant's related failure root cause.Clinical evaluation: revision surgery performed 3 years after cementless total hip arthroplasty in a (b)(6) man.Radiographic image provided shows a mobilized cup.Initial shell position cannot be determined as no immediate postoperative image is available.Aseptic loosening is a possible literature described adverse event after cementless total hip arthroplasty and causes are often unknown.The reason of this event cannot be determined.
|
| |
|
Event or Problem Description
|
|
Acetabular cup loosening unknown cause, due to mobilization there was an impingement cup- stem.Cup, head and liner were revised almost 3 years and 3 weeks after primary.
|
| |
|