Patient was revised to address loosening of the tibial component at cement to implant interface.Cement manufacturer is depuy.During the revision, the tray was found to be grossly loose and there was no cement adherence to the back side of the implant.The team attempted to revise the tibial cut and implant attune fb revision tibial tray.However, the original insert was 14mm thick and by the time the team freshened up the tibia cut , they were handed a 20mm thick attune insert because that was all they had.As a result, the team ended up taking off the femur and doing a full sigma knee revision.It was also noted that the original poly insert was severely pitted.Doi: (b)(6) 2017.Dor: (b)(6) 2018; right knee.
|