MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET Back to Search Results
Model Number 82400
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 12/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not find a conclusive cause for the higher-than-expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the trima accel system operated as intended by displaying the ¿verify platelet product, yield and volume¿ system advisory message due to the procedure being ended early.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related investigation is in process.A follow-up report will be provided.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit #: w036319315240 the platelet collection set is not available for return because it was discarded by the customer.
 
Additional Manufacturer Narrative
This report is being filed to provide additional information in h6 and h10.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Corrective action: an internal capa has been opened to evaluate white blood cell (wbc) contamination in relation to a recent increase in the number of reported wbc contaminations.Root cause: the analysis of the run data file did not find a conclusive cause for the higher-than- expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the trima accel system operated as intended by displaying the ¿verify platelet product, yield and volume¿ system advisory message due to the procedure being ended early.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key9602311
Report Number1722028-2020-00027
Device Sequence Number13542014
Product Code GKT
UDI-Device Identifier05020583824003
UDI-Public05020583824003
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
BK190332
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2021
Device Model Number82400
Device Catalogue Number82400
Device Lot Number1907311130
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2019
Supplement Date Received by Manufacturer04/06/2020
Initial Report FDA Received Date01/17/2020
Supplement Report FDA Received Date04/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-