| Catalog Number |
323487 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in (b)(4) has been listed and the fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.The customer's address is unknown.(b)(6) usa has been used as a default.Device manufacture date: unknown.Investigation summary: level a investigation complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_1__; occurrence: unable to perform complaint lot history check for difficult/unable to operate due to unknown lot number.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.If samples are received in the future the complaint will be reopened for further investigation.Unable to perform dhr check for difficult/unable to operate due to unknown lot number.Investigation conclusion: as no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that bd home sharps container would not allow the syringes to fit inside the container.This was discovered during use.The following information was provided by the initial reporter: material no.323487 batch no.Unknown.It was reported that syringes are unable to be disposed in sharps container.Verbatim: from phone call on 2020-01-02 16:15:55: returned call, parent stated her daughter uses the prefilled syringes and she is unable to dispose it in the sharp container.Explained the parent this container is not compatible for the prefilled syringes.
|
| |
|
Search Alerts/Recalls
|