| Catalog Number |
1550400-08 |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Migration (4003)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/03/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001.(b)(4).The device was not returned for analysis.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.It was reported the xience sierra stent delivery system (sds) was inflated above the rated burst pressure of 16 atms.It should be noted that the xience sierra everolimus eluting coronary stent system instruction for use states: do not exceed rated burst pressure (rbp) as indicated on product label.Monitor balloon pressure during inflation.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the balloon catheter used to perform post-dilation interacted with the deployed stent causing the reported stent migration with the treatments of additional therapy/non-surgical treatment, additional therapy/delayed treatment, and removal of a foreign body.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a non-tortuous, non-calcified left main lesion.The 4.0x08mm xience sierra stent was deployed at 20 atmospheres.Post-dilatation was done with a 5.5x20mm non-abbott balloon and when the balloon was retrieved, the stent came back with it and became loose in the anatomy.The stent was retrieved with a snare device.There was a 1 hour 10 minutes delay in the procedure.A new non-abbott stent was used to treat the lesion.There was no adverse patient sequela reported.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|