Model Number CLT-21-MG
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Codes Radiation Exposure, Unintended (3164); No Code Available (3191)
Date of Event 01/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during esophagus procedure, five pieces should be included, however, only four pieces were included at the time of needle counting.There were four packs, the event was noted when returning the needle at the 17th stitch, as the surgical field was not closed, both intra-abdominal ct and mri examination did not find any needle or thread.When the package sponge part was checked, only one place was found to be no needle was placed.The surgeon wondered whether the sponge part would close after taking the needle out and checked it for overnight, it was apparent that the sponge part had not been placed with another needle.The procedure was completed with another device.The surgical time was extended by more than 30 minutes.
 
Additional Manufacturer Narrative
Additional information: d10, g4, h3, h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted a retainer with no packaging that contained four needles and two wound sutures.Three remaining samples were received open with appropriate packaging.All five sutures and needles remained inside.It was reported that there was an incomplete needle quantity.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9607009
Report Number9612501-2020-00112
Device Sequence Number19500372
Product Code GAM
UDI-Device Identifier10884521044548
UDI-Public10884521044548
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberCLT-21-MG
Device Catalogue NumberCLT-21-MG
Device Lot NumberD9F0546Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2020
Initial Date Received by Manufacturer 01/07/2020
Supplement Date Received by Manufacturer02/26/2020
04/29/2020
Initial Report FDA Received Date01/20/2020
Supplement Report FDA Received Date03/12/2020
02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-