| Model Number |
SS085075M |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Embolus (1830)
|
| Date of Event |
12/20/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The physician confirmed the stroke was not related to the introducer sheath and there are no alleged deficiencies regarding the device.The results of the investigation are inconclusive as the device was not returned for analysis.A review of the device history record was not possible as the lot number was not reported.Per the ifu, embolism and stroke are associated risks with the use of this device.The ifu contains contraindications against use of this device for those with recent cerebral vascular accidents or embolic events.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the catheter introducer sheath the patient had a stroke while going transseptal.The physician was able to confirm that the stroke was not related to the introducer sheath and there are no alleged deficiencies regarding the device.
|
| |
|
Search Alerts/Recalls
|