| Model Number |
RONYX20015UX |
| Medical Device Problem Codes |
Activation, Positioning or Separation Problem (2906); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During a procedure a resolute onyx rx coronary drug-eluting stent was used to treat a moderately tortuous, moderately calcified lesion.There were no issues removing the device from the packaging.The device was inspected with no issues noted.Negative prep was performed with no issues.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered and excessive force was not used.It was reported that the stent was unable to be delivered to the lesion with it being removed and attempted multiple times, it was inspected again after removal before each attempt.It was noted that the stent was not able to be delivered to the lesion due to difficult anatomy.On the final attempt the stent did cross the lesion and the balloon was inflated to deploy the stent.There were no inflation difficulties.It was reported that upon its removal the stent was still on the balloon and was noted to be severely deformed.It is stated that it was thought the tuohy adapter was too tight and the stent 'accordianed' there.It is also stated that it was thought not to be the stent's fault.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the device returned with a non-medtronic stop-cock attached to the luer.The stent was not positioned on the balloon between the marker bands.It was severely bunched and deformed, and was located over the proximal marker-band.Stent measurements could not be recorded due to the nature of deformation.Balloon folds were expanded with contrast visible in between balloon folds.A kink was evident to the distal shaft, with the underlying inner member and support wire also kinked.Deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|