It was reported that the valve was connected in the patient to the ventricle and abdominal catheter.It was then found that the valve was damaged/broken and leaking.As a result, the surgery could not be carried out, and all the devices were removed.It was stated that there was a procedure delay of 60 minutes.The skull was closed, and they were awaiting back up products for surgery again.The patient's status at the time of the report was alive-no injury.Additional information received stated that the patient had an additional surgery to implant another device on a different date.It was indicated that there were no environmental/external/patient factors that may have led or contributed to the issue.
|
|
The returned valve was patent.It met the requirements for valve flux, and siphon control and preimplantation testing.The catheter from pe 703534082-20 and 703534082-30 were received with the valve.The valve did not meet the requirements for leakage, reflux and pressure flow testing due to tear on the delta chamber.This tear was not visually evident but air bubbles around the delta chamber were observed during leakage testing.The instruction for use cautions, ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components.Such damage may lead to loss of shunt integrity¿¿ dhr review of lot # e45625 did not indicate any findings that correlate to the complaint.All valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|