• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY VENTRICULAR CATHETER, STANDARD, BARIUM IMPREGNATED, 23 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY VENTRICULAR CATHETER, STANDARD, BARIUM IMPREGNATED, 23 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 41101
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the valve was connected in the patient to the ventricle and abdominal catheter.It was then found that the valve was damaged/broken and leaking.As a result, the surgery could not be carried out, and all the devices were removed.It was stated that there was a procedure delay of 60 minutes.The skull was closed, and they were awaiting back up products for surgery again.The patient's status at the time of the report was alive-no injury.Additional information received stated that the patient had an additional surgery to implant another device on a different date.It was indicated that there were no environmental/external/patient factors that may have led or contributed to the issue.
 
Additional Manufacturer Narrative
The returned product met the requirements for patency and leak testing.The catheter was packaged with valve from pe (b)(4).Review of dhr for lot # e58478 did not indicate any findings that correlate to the complaint.All catheters are 100% inspected at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRICULAR CATHETER, STANDARD, BARIUM IMPREGNATED, 23 CM
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key9611797
Report Number2021898-2020-00020
Device Sequence Number523736
Product Code JXG
UDI-Device Identifier00643169466500
UDI-Public00643169466500
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K792007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number41101
Device Catalogue Number41101
Device Lot NumberE58478
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/25/2019
Supplement Date Received by Manufacturer03/20/2020
Initial Report FDA Received Date01/21/2020
Supplement Report FDA Received Date03/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-