| Model Number |
D003 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
12/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded code 1005 indicative shorted shock on the right ventricular (rv) lead.Intervention was performed.The device was explanted and the right ventricular (rv) lead was capped.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Patient code 3191 captures the reportable event of additional surgical intervention.
|
| |
|