| Model Number |
N/A |
| Medical Device Problem Codes |
High Test Results (2457); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cause for the discordant ca19-9 results is unknown.Siemens healthcare diagnostics is investigating.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the ifu states in the limitations section: warning "do not use the advia centaur ca 19-9 assay as a screening test or for diagnosis.Do not predict disease recurrence solely on levels of advia centaur ca 19-9.Normal levels of advia centaur ca 19-9 do not always preclude the presence of disease.Note do not interpret serum levels of ca 19-9 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed gi carcinoma frequently have levels of ca 19-9 within the range observed in healthy individuals.Additionally, elevated levels of ca 19-9 can be observed in patients with nonmalignant diseases.Measurements of ca 19-9 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.The concentration of ca 19-9 in a given specimen determined with assays from different manufacturers can vary because of differences in assay methods, calibration, and reagent specificity.Ca 19-9 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.Therefore, it is important to use assay specific values to evaluate quality control results." mdr 1219913-2020-00023 was filed for the same event for advia centaur xp #1 result.
|
| |
|
Event or Problem Description
|
|
Two discordant high results were obtained on one patient sample with the advia centaur xpt ca19-9 immunoassay.The patient sample was tested on two advia centaur xp systems and similar high results were obtained between the systems.The results were reported to physician and questioned.A new sample from the same patient was obtained at a later date and tested with atellica im ca19-9 immunoassay on two systems.Lower results were obtained on both systems.A corrected report was issued with the lower results.Quality control was within specification for all systems used for testing.Patient treatment was not prescribed or altered.There was no report of adverse health consequences due to the discordant high ca19-9 results.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics filed initial mdr 1219913-2020-00024 on 01/21/2020.Additional information from 04/02/2020: siemens healthcare diagnostics has investigated a customer complaint of falsely elevated results on one patient sample using the advia centaur xp ca 19-9 immunoassay compared to results from a redraw of the same patient using the atellica im ca 19-9 immunoassay at a later date.Requests for the samples were made by siemens for further investigation, however the samples were not available.Without the sample, siemens is unable to determine whether there was a true change in patient ca19-9 levels or if a platform or reagent lot difference was the root cause of the discordancy.Siemens reviewed manufacturing quality control results from advia centaur ca19-9 reagent lot 452 (complaint lot) and atellica im ca19-9 reagent lot 456 and all specifications were met.In addition, routine troubleshooting did not identify an issue in the customer's laboratory.Siemens concluded that there is a potential specific issue with the patient sample, and possibly a change in patient treatment between the draw dates.Pre-analytical variables can affect the quality of the sample, and deviation from recommended best practices can impact results.While there is insufficient information to determine the cause of the discordant result, siemens cannot rule out pre-analytical variables or sample specific issues as contributing factors.No product problem has been identified.No further investigation required.Section h6, the device, method, results and conclusion codes were updated to reflect the additional information.Mdr 1219913-2020-00023 supplemental report 1 was filed for the same event.
|
| |
|
Search Alerts/Recalls
|