• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA25-80/I16-40
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.If additional information is obtained at a later time that is pertinent to this event, a follow-up report will then be submitted.Device iteration is afx with duraply.Devices remain implanted in patient.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with the afx abdominal aortic aneurysm stent.Approximately five (5) years post initial procedure (exact date was not provided), the patient presented at routine follow-up with aneurysm sac growth (10mm) and an endoleak of indeterminate origin.The physician plans to re-intervene and will continue to monitor the patient.The patient is reportedly asymptomatic.As of date, there have been no additional patient sequelae reported.
 
Additional Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.A clinical assessment of the reported event could not be completed due to a lack of receipt of relevant medical records and/or imaging.Several attempts were made to obtain medical records and medical images but was denied as patient authorization is needed.As such, procedure related harms, event determination, off label conditions, related patient harms and patient disposition could not be assessed.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Device iteration is afx with duraply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
MDR Report Key9613342
Report Number2031527-2020-00023
Device Sequence Number2677449
Product Code MIH
UDI-Device Identifier00818009010018
UDI-Public(01)00818009010018(17)170703
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/03/2017
Device Model NumberBA25-80/I16-40
Device Lot Number1251741-023
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/26/2019
Supplement Date Received by Manufacturer12/26/2019
Initial Report FDA Received Date01/21/2020
Supplement Report FDA Received Date07/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL 1264514-013
Outcome Attributed to Adverse Event Other;
Patient Age87 YR
-
-