• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TROCHANTERIC STABILIZING PLATE OMEGA 3 135°, 4 HOLES; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TROCHANTERIC STABILIZING PLATE OMEGA 3 135°, 4 HOLES; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 597404S
Medical Device Problem Code Difficult or Delayed Positioning (1157)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was not returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition unknown.
 
Event or Problem Description
As reported: "the omega 3 system was chosen to provide fixation of an intertrochanteric fracture on a male trauma patient of unknown age.The omega 135° guide was used to insert the omega guide pin.Omega reamer and omega tap were used before insertion of lag screw.An omega 3 lag screw (95mm) was implanted over the omega guide pin.The surgeon decided to exchange it for a longer lag screw.The rear part (screw/ driver interface) of the 95mm lag screw was visually deformed after extraction.An omega 3 lag screw (105mm) was then inserted over the guide wire.During insertion, the distal (rear) part of the screw began to deform (at the screw/ driver interface).The deformation was visualized by the surgeon.The surgeon attempted to install the omega 3 stabilization plate onto the lag screw without success.The lag screw screwdriver was not able to remove the damaged lag screw.A conical extractor was used to successfully remove the lag screw.A new omega 3 lag screw (100mm) was then inserted over the guide wire.The omega stabilization plate did not fit the bone to the surgeon's liking and was replaced with an omega short barrel hip plate 135°, 2 hole.X (4.5 x 40mm) screws were inserted through the plate into the femur shaft to secure the plate to the bone." additionally reported: "surgery was prolonged about an hour and a half because of the implant malfunction.".
 
Event or Problem Description
As reported: "the omega 3 system was chosen to provide fixation of an intertrochanteric fracture on a male trauma patient of unknown age.The omega 135° guide was used to insert the omega guide pin.Omega reamer and omega tap were used before insertion of lag screw.An omega 3 lag screw (95mm) was implanted over the omega guide pin.The surgeon decided to exchange it for a longer lag screw.The rear part (screw/ driver interface) of the 95mm lag screw was visually deformed after extraction.An omega 3 lag screw (105mm) was then inserted over the guide wire.During insertion, the distal (rear) part of the screw began to deform (at the screw/ driver interface).The deformation was visualized by the surgeon.The surgeon attempted to install the omega 3 stabilization plate onto the lag screw without success.The lag screw screwdriver was not able to remove the damaged lag screw.A conical extractor was used to successfully remove the lag screw.A new omega 3 lag screw (100mm) was then inserted over the guide wire.The omega stabilization plate did not fit the bone to the surgeon's liking and was replaced with an omega short barrel hip plate 135°, 2 hole.X (4.5 x 40mm) screws were inserted through the plate into the femur shaft to secure the plate to the bone." additionally reported: "surgery was prolonged about an hour and a half because of the implant malfunction.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC STABILIZING PLATE OMEGA 3 135°, 4 HOLES
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key9614101
Report Number0008031020-2020-00115
Device Sequence Number2922772
Product Code KTT
UDI-Device Identifier07613154687428
UDI-Public07613154687428
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K081278
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model Number597404S
Device Catalogue Number597404S
Device Lot NumberD30505
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/26/2019
Supplement Date Received by Manufacturer01/30/2020
Initial Report FDA Received Date01/21/2020
Supplement Report FDA Received Date02/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-