| Model Number |
657312S |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
When inserting the locking screw (657312s) into the plate ((b)(4)), a thin metal piece was generated.It has been removed and no longer remains in the body.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed.The device inspection revealed the following: the screw as well as some metal abrasion / swarf was returned.The damage of the screw thread is clearly visible, the metal abrasion has come of the entire middle section of the screw thread.A close up (taken by the microscope) clearly indicates that the screw had been inserted on an angle from the start, as the metal swarf was getting from thin to thick as the screw was inserted further till the final breakage clearly proves that as well.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.The failure was caused by an inadequate applied angle during screw insertion.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
When inserting the locking screw (657312s) into the plate (4020903s), a thin metal piece was generated.It has been removed and no longer remains in the body.
|
| |
|