In (b)(6) 2019, the patient noted a protrusion and some fluid leaking around a portion of the rns system incision site.The patient was scheduled for a revision procedure on (b)(6) 2020.During the procedure, it was determined that one of the unconnected, capped leads was applying outward pressure near the incision line, preventing proper healing.The neurosurgeon re-positioned the lead, securing it away from the incision site.No further information was provided by the treating center.
|